EMB shows UV absorption only in the low wavelengthsof the UV spectrum  translation - EMB shows UV absorption only in the low wavelengthsof the UV spectrum  Indonesian how to say

EMB shows UV absorption only in the

EMB shows UV absorption only in the low wavelengths
of the UV spectrum due to the absence of chromophores in
its structure. Thus, analyses were performed at 200, 205, and
210 nm. It was observed that at very low wavelengths (200 and
205 nm), baseline fluctuation and signal noise were significantly
higher, so that a wavelength of 210 nm was chosen for EMB
analysis.
RIF, INH, and PYZ showed high signals at several
wavelengths; therefore, 238, 254, 265, and 338 nm were tested.
Adequate results were observed at 238 nm for the three drugs.
Consequently, simultaneous quantification was possible using
a wavelength of 210 nm for EMB and 238 nm for RIF, INH,
and PYZ.
0/5000
From: -
To: -
Results (Indonesian) 1: [Copy]
Copied!
EMB menunjukkan UV penyerapan hanya dalam panjang gelombang rendahspektrum UV karena tidak adanya chromophores distrukturnya. Dengan demikian, analisis dilakukan di 200, 205, dan210 nm. Terlihat bahwa di panjang gelombang yang sangat rendah (200 dan205 nm), dasar fluktuasi dan sinyal kebisingan secara signifikanlebih tinggi, sehingga gelombang 210 nm dipilih untuk EMBanalisis.RIF, INH dan PYZ menunjukkan sinyal tinggi di beberapapanjang gelombang; oleh karena itu, 238, 254, 265 dan 338 nm diuji.Hasil yang memadai diamati di 238 nm untuk obat-obatan tiga.Akibatnya, simultan kuantifikasi ini mungkin menggunakangelombang 210 nm EMB dan 238 nm untuk RIF, INH,dan PYZ.
Being translated, please wait..
Results (Indonesian) 2:[Copy]
Copied!
EMB menunjukkan penyerapan UV hanya dalam panjang gelombang rendah
dari spektrum UV karena tidak adanya kromofor di
strukturnya. Dengan demikian, analisis dilakukan pada 200, 205, dan
210 nm. Diamati bahwa pada panjang gelombang yang sangat rendah (200 dan
205 nm), fluktuasi baseline dan sinyal noise secara signifikan
lebih tinggi, sehingga panjang gelombang 210 nm dipilih untuk EMB
analisis.
RIF, INH, dan PYZ menunjukkan sinyal tinggi di beberapa
panjang gelombang; Oleh karena itu, 238, 254, 265, dan 338 nm diuji.
Hasil yang memadai diamati pada 238 nm untuk tiga obat.
Akibatnya, kuantifikasi simultan adalah mungkin menggunakan
panjang gelombang 210 nm untuk EMB dan 238 nm untuk RIF, INH,
dan PYZ .
Being translated, please wait..
 
Other languages
The translation tool support: Afrikaans, Albanian, Amharic, Arabic, Armenian, Azerbaijani, Basque, Belarusian, Bengali, Bosnian, Bulgarian, Catalan, Cebuano, Chichewa, Chinese, Chinese Traditional, Corsican, Croatian, Czech, Danish, Detect language, Dutch, English, Esperanto, Estonian, Filipino, Finnish, French, Frisian, Galician, Georgian, German, Greek, Gujarati, Haitian Creole, Hausa, Hawaiian, Hebrew, Hindi, Hmong, Hungarian, Icelandic, Igbo, Indonesian, Irish, Italian, Japanese, Javanese, Kannada, Kazakh, Khmer, Kinyarwanda, Klingon, Korean, Kurdish (Kurmanji), Kyrgyz, Lao, Latin, Latvian, Lithuanian, Luxembourgish, Macedonian, Malagasy, Malay, Malayalam, Maltese, Maori, Marathi, Mongolian, Myanmar (Burmese), Nepali, Norwegian, Odia (Oriya), Pashto, Persian, Polish, Portuguese, Punjabi, Romanian, Russian, Samoan, Scots Gaelic, Serbian, Sesotho, Shona, Sindhi, Sinhala, Slovak, Slovenian, Somali, Spanish, Sundanese, Swahili, Swedish, Tajik, Tamil, Tatar, Telugu, Thai, Turkish, Turkmen, Ukrainian, Urdu, Uyghur, Uzbek, Vietnamese, Welsh, Xhosa, Yiddish, Yoruba, Zulu, Language translation.